Spectrum Master™ MS2723C Specifications
MS2723C TDS PN: 11410-00524 Rev. J 5
High Accuracy Power Meter (option 19, Requires external USB Power Sensor)
Setup Parameters
Amplitude
Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average
# of Running Averages, Max Hold
Zero/Cal
Zero On/Off, Cal Factor (Center Frequency, Signal Standard)
Limits
Limit On/Off, Limit Upper/Lower
Power Sensor Model PSN50 MA24104A/105A MA24106A MA24108/18/26A
Description
High Accuracy
RF Power Sensor
Inline High/Peak
Power Sensor
High Accuracy
RF Power Sensor
Microwave
USB Power Sensor
Frequency Range
50MHz to 6GHz 600MHz to 4GHz
(MA24104A)
350MHz to 4GHz
(MA24105A)
50 MHz to 6 GHz 10 MHz to 8 GHz
(MA24108A)
10 MHz to 18 GHz
(MA24118A)
10 MHz to 26 GHz
(MA24126A)
Connector
Type N (m), 50
Ω
Type N(f ), 50
Ω
(MA24104A)
Type N(f ), 50
Ω
(MA24105A)
Typ e N ( m ), 50
Ω
Type N(m), 50
Ω
(MA24108/18A)
Typ e K ( m ) , 5 0
Ω
(MA24126A)
Dynamic Range
–30dBm to +20dBm
(0.001 mW to 100 mW)
+3 dBm to +51.76 dBm
(2 mW to 150 W)
–40dBm to +23dBm
(0.1 µW to 200 mW)
–40dBm to +20dBm
(0.1 µW to 100 mW)
VBW 100 Hz 100 Hz 100 Hz 50 kHz
Measurands
True-RMS True-RMS True-RMS True-RMS, Slot Power,
Burst Average Power
Measurement
Uncertainty
±0.16 dB
1
±0.17 dB
2
±0.16 dB
1
±0.18 dB
3
Datasheet
(for complete
specifications)
11410-00414 11410-00483
(MA24104A)
11410-00621
(MA24105A)
11410-00424 11410-00504
Notes:
1. Total RSS measurement uncertainty (0 °C to 50 °C) for power measurements of a CW signal greater than
–20 dBm with zero mismatch errors.
2. Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a
matched load. Measurement results referenced to the input side of the sensor.
3. Expanded uncertainty with K=2 for power measurements of a CW signal greater than –20 dBm with zero
mismatch errors.